impact instruments 25

 

 + 27 (0) 11 704 7001 / 5 / 8

 info@impactinstruments.co.za

Bolt Tension Meter

TI-MINIMAX / BT1-DL Ultrasonic Bolt Tension Monitor

The MINIMAX (now known as the BT1-DL) is an Ultrasonic Bolt Tension Monitor designed to measure elongation, time, load, stress and percentage strain.This eliminates issues from traditional torque methods by directly measuring tension.

bolt tension illustration 

           Bolt Tension Meter   Bolt Tension Meter 2   

Features

 
Model BT1-DL

 

Display Mode

Material thickness digits display
 •
A-Scan
Rectified +/- (half wave view), or RF (full waveform view)
Measurement Range
PE - 25.4 - 25,400mm (1 - 999”)
Resolution
0.001mm (0.0001”) or 0.0001mm (0.00001”) switchable
Measurement Rate
Manual
8 readings per second
Velocity Calibration Range
1250 - 13,995m/s (0.0492 - 0.5510 in/μs)
Temperature Compensation
 •
Limit Alarm Mode
 •
Memory and Data Logging
4GB internal memory
Sequential and grid batching
Bitmap graphic capture and capture viewer
Calibration Options
Automatic, fixed, single and two-point
Transducer Probe Type
Single element
Transducer Frequency Range
1 - 10MHz
Display
1/8” VGA greyscale display (240 x 160 pixels)
Viewable area 62 x 45.7mm (2.4 x 1.8”)
Units (Selectable)
in/mm, μs, °F/°C
LED Backlight
On / Off/ Auto
Repeatability / Stability Indicator
 •
Low Battery Indicator
 •
Battery Save Mode
Auto
Transducer Connector Type
Microdot, Lemo and BNC
USB-C Interface
 •
Temperature Sensor
5 pin LEMO 1
Screen Refresh Rate
30 Hz
Timing
Precision TCXO timing with single shot 100 MHz 8 bit ultra-low power digitizer
Pulser Type
Square Wave. Selectable 100, 150 and 200 volts. Selectable options Spike, Thin, and Wide. 80 to 400ns
Damping
50, 75, 100, 300, 600, & 1500ohms
Frequency Band
Broadband 1.8 - 19 MHz (-3dB) filter
Horizontal Linearity
+/- 0.4% FSW
Vertical Linearity
+/- 1% FSH
Amplifier Linearity
+/- 1 dB
Amplitude Measurement
0 to 100% FSH, with 1% resolution
Delay
25,400 mm (0 - 999.9”) at steel velocity
Measurement Gate
One gate with audible and visual alarm. Amplitude 5-95%, 1% steps
Setups
64 custom user defined setups; factory setups can also be edited by the user
Auto Set
Automates the detection, scope, and display setting process for each individual bolt
Alarm Limits
Adjustable Hi/Low tolerances with visual LED’s and audible beeper
Field Calibration
Vector or Regression correction curve for increased accuracy using Load & Stress
Alarm Limits Bar
Hi & Lo alarm limits for displaying an acceptable tolerance range
Detections
Zero Crossing
Time
Nanoseconds
Elongation
Change in length (in/mm)
Load
Force load applied (pounds KIP, or megapascals MPa)
Stress
Force for unit area stress applied (inches per inch or millimeters per millimeter)
%Strain
Ratio of elongation to effective length
Bolt Materials
Select types from a preset or custom list
Glue-On
Available for short bolts with minimal/short elongations to eliminate transducer placement errors
Operating Temperature                                                 -10 to 60ºC (14 to 140ºF)
Power Supply 3 x AA batteries and via USB
Battery Life Alkaline 35 hrs, Nicad 10 hrs, NI-MH 35 hrs
Gauge Weight 383g (13.5oz) - including batteries
Gauge Dimensions 63.5 x 165 x 31.5mm (2.5 x 6.5 x 1.24”)
Packing List Unit, 3 AA Batteries, 4 oz. Bottle of Couplant, USB Type A to Type C Transfer Cable, Carrying Case, 10’ Lemo to Microdot Transducer cable, PC Utility Software (download from website), Operation Manual and Certification.

 

 Probes/Transducers and Calibration blocks are sold separately

pdficon largeBrochure

Our Worldwide Partners

Alvib Sistemas Checkline Logo 2023 cicor DMQ logoGuide Sensmart LogoUltrasonic Flow Logo ion science logo new Kohtect logo new Madur logo PCE Logo Prosense logo 2021Seiffert Industrial Seitron Logo Semapi Logo Shimpo Instruments Logo SpotSee New Logo full Technoac Vibrasens logo